beautypg.com

Analog scaling, Analog calibration – APG IRU-9400 user manual User Manual

Page 26

background image


IRU Series

Rev. B1 3/10

26

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

• Analog Scaling

To scale the analog output, simply enter the desired distance, beginning at the

sensor face, to each of the two end points of the analog signal using the 4 mA/0

V Set Point

and 20 mA/2.5/5 V Set Point parameters. The analog output will be

scaled between the two distance values entered in these settings. The output can

be scaled in either direction; signal rising or falling as the distance increases (see

diagram below). If the detected target is outside the scaled range, the sensor will

output either full scale or minimum scale depending on which end of the analog

scale the target resides.

• Analog Calibration

(No adjustments required in most applications)

Calibration of the analog signal is accomplished using the 4 mA Calibration

and 20 mA Calibration or the 2.5/5 Volt Calibration parameters. To calibrate

the output, the sensor must be outputting the analog endpoint value that you

wish to calibrate (i.e, full scale or minimum scale). Increasing or decreasing the

value in the calibration parameter will increase or decrease the corresponding

output signal value. For example, decreasing the 20 mA Calibration value will

decrease the 20 mA signal output value. These parameters are factory set and

normally should not require adjustment. The default values (shown below) will

bring the output to within a few mirco-amps or micro-volts of an exact analog

signal value.

4 mA Calibrate = 8200

20 mA Calibrate = 41000

2.5 Volt or 5 Volt Calibrate = 4095

4 mA

20 mA

20 mA

4 mA

Increasing Distance

0 V

2.5/5 V

2.5/5 V

0 V

This manual is related to the following products: