beautypg.com

APG IRU-9400 user manual User Manual

Page 23

background image

Rev. B1, 3/10

IRU Series

23

Automation Products Group, Inc.

APG...Providing tailored solutions for measurement applications

Tel: 1/888/525-7300 • Fax: 1/435/753-7490 • www.apgsensors.com • [email protected]

Sample Rate:

Sets how many times per second the sensor takes a reading.

The Sample Rate is expressed in Hertz. Options allow rates from 1 to 50 Hz.

The maximum Sample Rate setting is determined by the Max Distance setting.

The greater the Maximum Distance setting, the slower the maximum allowable

Sample Rate setting. A higher Sample Rate will achieve a more rapid response

to target movements but can also decrease the stability of the measurement.

Lowering the Sample Rate will help increase sensor life and decrease the chance

of detecting errant echoes. It is highly recommended that the Sample Rate be

set only as high as is necessary for the application. A sample rate of 1-2 Hz is

usually appropriate for slow moving level sensing applications.

Multiplier:

Used to true the sensor’s distance reading with the actual physical

measurement to the target. Since the speed of sound is not constant through

all environments and temperatures, the multiplier parameter allows the user

to adjust for these variations in atmosphere. To determine the appropriate

multiplier setting, measure from the face of the sensor to the intended target.

Then divide the measured distance by the distance reading of the sensor, and

enter the resulting value in the Multiplier setting. For example, if the target is

measured at 2.5 ft. from the sensors face, and the software distance reading is

2.61 ft. The multiplier setting would be 2.5 / 2.61, which = 0.9579

Offset:

Used to adjust the zero reference point of the sensor. When the Offset

is set to 0, the zero reference of the sensor is at the face of the transducer.

Setting the Offset to a negative number will move the zero reference backward

(behind the sensor face), while a positive setting will move the reference

forward (in front of the sensor face).

Temp_Comp:

Used to enable or disable the internal temperature

compensation of the sensor. The speed of sound through air changes about

0.18% for every 1

0

C of temperature change. Enabling the internal temperature

compensation can reduce the effects of temperature changes by 50% or more,

depending on the temperature gradient through the sensing range.

This manual is related to the following products: