beautypg.com

10 technical data, 1 specifications, Specifications – Eppendorf Galaxy Gas Analyzer User Manual

Page 45: Galaxy® gas analyzer — operating manual, Technical data

background image

45

Galaxy® Gas Analyzer — Operating manual

10

Technical data

10

Technical data

10.1

Specifications

Galaxy Gas Analyzer Specifications

Mains/Power Supply
Battery Type

Lithium ion

Battery Life before recharge

10 hours (8 hours with pump working)

Battery Lifetime

> 300 cycles

Battery Charger

5 V DC external mains/power supply and internal charging circuit

Charge Time

3 hours

Alternative Power

USB connector
Mains/DC power supply

Gases
Gases Measured

CO

2

By custom dual wavelength infrared cell with reference
channel

O

2

(optional)

By internal electrochemical cell

Oxygen Cell Lifetime

Approximately 3 years in air environment

Range

CO

2

0 to 20 %

O

2

(optional)

0 to100 %

Measurement Accuracy*

CO

2

Accurate to ± 1 % range after calibration

O

2

(optional)

Accurate to ± 1 % range after calibration

Response Time, T

90

CO

2

 20 seconds

O

2

(optional)

 60 seconds

Other Features
Temperature (optional)

Two temperature probes available, range 0 °C to +50 °C

Temperature Accuracy

Typically, ± 0.1 °C from 32 to 44 °C, ± 0.2 °C over the remaining range

Barometric Pressure

800 to 1200 mbar

RH Measurement (opt.)

RH probe available, 0 to 100 % RH non-condensing

RH Accuracy

± 1.5 % RH across the range

Visual & Audible Alarm

User-selectable CO

2

and O

2

alarm levels

Communications

USB type B mini-connector, HID device class

Data Storage

1000 reading sets + 270 events

Pump
Flow

Typically, 100 cc/min

Environmental Conditions
Operating Temperature Range

0 °C to 50 °C

10

Technical data