8 maintenance, 1 introduction, 2 replacement parts – Eppendorf Galaxy Gas Analyzer User Manual
Page 38: 3 cleaning, Maintenance 8.1, Introduction, Replacement parts, Cleaning, Galaxy® gas analyzer — operating manual, Maintenance

Galaxy® Gas Analyzer — Operating manual
38
8
Maintenance
8
Maintenance
8.1
Introduction
The Galaxy Gas Analyzer should be regularly serviced to ensure correct operation and accurate
readings.
Depending on usage, the O
2
cell should be replaced by the manufacturer every 2 - 3 years.
8.2
Replacement parts
The following parts are supplied by the manufacturer for your use with the Galaxy Gas Analyzer:
8.3
Cleaning
The instrument and accessories (including battery charger) can be wiped clean using a
non-fibrous damp cloth.
NOTICE!
Risk of material damage
There are no user-servicable parts inside the instrument. Do not attempt to open the gas
analyzer, as this may invalidate the warranty.
Part
Used
Sample Filter
Regularly inspect the sample filter, and replace it when you
find signs of damage or discoloration.
Never operate the instrument without the sample filter, as this
may result in water or dust entering the gas analyzer.
Replace the filter immediately if water is seen; failure to do so
can damage the instrument.
Sample Tubing
Always inspect the sample tubes to be sure they are not
contaminated or damaged. Replace them as needed.
NOTICE!
Risk of material damage
Do not apply pressure to the LCD display as this can cause damage.
Do not apply any moisture to the moisture removal tube on the rear of the instrument as this
may damage.
Do not use solvents or any other chemical cleaners.
8
Maintenance