Galaxy® Gas Analyzer — Operating manual

44

Abb. 29:

Typical Zero Gas Purge Time

Fig. 29:

Typical Zero Gas Purge Time

9.4

No cross-gas effects

Carbon dioxide is measured by infrared absorption at a wavelength specific to that gas.
Therefore, the carbon dioxide reading will not be affected by any other gases.
The oxygen sensor is a galvanic cell type and experiences virtually no influence from CO

2

, CO,

H

2

S, NO

2

, SO

2

or H

2

, unlike many other types of oxygen cells.

9.5

Hardware reset

If for any reason the instrument locks up and will not turn off, it is possible to force a hardware
reset: press and hold the On/Off key for 10 seconds; wait at least 15 seconds, and the instrument
should restart automatically.

Hint!

Performing a hardware reset may cause loss or corruption of currently stored data, including the
time/date. If this happens, be sure to reset the time and date.

9

Troubleshooting