Eppendorf I-26R Large Stackable Incubator Shakers User Manual
Page 31
31
New Brunswick
Operating Manual
8.2
Continuous (Unlimited) Run
1. Press
SELECT
until the
RPM INDICATOR
is illuminated.
2. If the display indicates that the shaker is
OFF
, press the
START/STOP KEY
.
3. Press either
▲(UP), ▼(DOWN)
KEY
to enter
SET MODE
(the
SET INDICATOR
will
illuminate).
4. Set the speed by using the
▲
or
▼
KEY
until the desired setpoint is displayed.
Continued pressure on the
▲
or
▼
KEY
will cause the setting to change more rapidly.
The setpoint may be changed during a run without stopping the shaker by
following steps 1-4 above.
8.3
Checking Any Setpoint
1. Press
SELECT
until the desired indicator is illuminated.
2. Briefly press either the
▲
or
▼ KEY
to enter the
SET MODE
and display the current
setpoint.
Holding the ▲ or ▼ for more than 0.5 seconds causes the speed
setpoint to change. Should this occur, resetting will be necessary.
8.4 Timed
Functions
The shaker may be programmed to automatically stop after a preset time period of 0.1 to
99.9 hours. There must be power to the shaker in order to set the timer, although a timed
run can be initiated while the unit is either stopped or operating.
To set the timer:
1. Press the
SELECT KEY
until the
HRS INDICATOR
is illuminated.
2. Press either
▲
or
▼ KEY
to enter the
SET MODE
and set the desired run time, between
0.1 and 99.9 hours.
If the shaker is stopped, skip to Step 5 below. If the shaker is already running:
3. Press the
START/STOP KEY
. The shaker will stop and the display will read
OFF
.
4. Press the
START/STOP KEY
again; the
TIME INDICATOR
will light and the shaker will
start the timed run.
If the shaker is stopped:
5. Press the
START/STOP KEY
. The shaker will start in untimed mode.