5 decontamination before shipment, 9 technical data, 2 mastercycler, thermomixer and thermostat – Eppendorf Temperature Verification System User Manual
Page 34: 3 maximum permissible errors and accuracy

Temperature Verification System USB – Single-channel — Operating manual
34
1. Switch off the Temperature Verification System USB – Single-channel and remove it from the
mains/power supply.
2. Let the Temperature sensor cool off.
3. Select a disinfection method that complies with the legal requirements and regulations in
place for your range of application.
4. Spray the Temperature sensor with a disinfectant. After the exposure time of the disinfectant
has elapsed, wipe off the sensor using a damp, lint-free cloth.
8.5 Decontamination before shipment
If you are shipping the device to the authorized Technical Service for repairs or to your authorized
dealer for disposal please note the following:
9
Technical data
9 Technical data
9.1 Temperature Verification System USB – Single-channel
The technical data can be found in the operating manual of the measuring device.
9.2 Mastercycler, Thermomixer and ThermoStat
The technical data is available in the operating manual of the devices you are verifying or
adjusting.
9.3 Maximum permissible errors and accuracy
9.3.1
Temperature Verification System USB – Single-channel
WARNING!
WARNING! Risk to health from contaminated device
1. Observe the information in the decontamination certificate, You find it as
a PDF file on our website (www.eppendorf.com/decontamination).
2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.
Temperature Verification System USB
– Single-channel
Measuring range
Accuracy
Measuring device with grip and sensor
35 °C – 95 °C
±0,3 K