Temperature Verification System USB – Single-channel — Operating manual

34

1. Switch off the Temperature Verification System USB – Single-channel and remove it from the

mains/power supply.

2. Let the Temperature sensor cool off.
3. Select a disinfection method that complies with the legal requirements and regulations in

place for your range of application.

4. Spray the Temperature sensor with a disinfectant. After the exposure time of the disinfectant

has elapsed, wipe off the sensor using a damp, lint-free cloth.

8.5 Decontamination before shipment

If you are shipping the device to the authorized Technical Service for repairs or to your authorized

dealer for disposal please note the following:

9

Technical data

9 Technical data

9.1 Temperature Verification System USB – Single-channel

The technical data can be found in the operating manual of the measuring device.

9.2 Mastercycler, Thermomixer and ThermoStat

The technical data is available in the operating manual of the devices you are verifying or

adjusting.

9.3 Maximum permissible errors and accuracy

9.3.1

Temperature Verification System USB – Single-channel

WARNING!

WARNING! Risk to health from contaminated device

1. Observe the information in the decontamination certificate, You find it as

a PDF file on our website (www.eppendorf.com/decontamination).

2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.

Temperature Verification System USB

– Single-channel

Measuring range

Accuracy

Measuring device with grip and sensor

35 °C – 95 °C

±0,3 K