Temperature Verification System USB – Single-channel — Operating manual
24
3. In the navigation tree, open the node of the cycler to be verified.
4. Select the
System
menu item.
5. Select the
Verification
tab.
6. Start the verification using the icon.
7. Select the Temperature Verification System USB – Single-channel using the checkbox.
8. Use the icon to confirm the selection.
You will be prompted to connect the Temperature Verification System USB – Single-channel.
9. Connect the Temperature Verification System USB – Single-channel.
10. Switch on the Measuring device.
The temperature sensor will be checked by the Mastercycler.
If the certification of the temperature sensor is no longer valid, an error message will be
shown. End the verification. Arrange to have the temperature sensor certified.
11. Confirm the query on the screen using the icon.
The verification program starts and heats the heated lid to 105 °C. You will then be prompted
to insert the temperature sensor (
Insert sensor in the position indicated
).
12. Open the heated lid.
13. Insert the temperature sensor in the position indicated.
14. Close the heated lid.
If a motor-driven heated lid is mounted, it automatically moves to the front position during the
next step.
15. Start the measuring cycle using the icon.
The measuring procedure runs at 95 °C and then at 35 °C.Each additional measuring cycle
begins with the request to move the temperature sensor to a different position.
16. Start each new measuring cycle using the icon.
The protocol will be shown on the monitor after the verification.
17. Complete the verification procedure using the icon.