3 adjusting the mastercycler ep – Eppendorf Temperature Verification System User Manual

Page 27

background image

Temperature Verification System USB – Single-channel — Operating manual

27

5.2.3

Adjusting the Mastercycler ep

Prerequisites

The Mastercycler is switched on.

The Measuring device is connected to the temperature sensor and switched on.

1. Log in as

Administrator

.

2. Select the

System

menu item under the cycler node.

3. Press the

Adjustment

softkey.

A note on the Temperature Verification System USB – Single-channel will be shown.

4. Confirm the note with the

OK

softkey.

You will be asked if you would like to use the Temperature Verification System USB –

Single-channel.

5. Answer the question using the

Yes

softkey.

You will be prompted to connect the Temperature Verification System USB – Single-channel.

6. Connect and switch on the Temperature Verification System USB – Single-channel.

The connection of the Temperature Verification System USB – Single-channel to the

Mastercycler devices is described in the installation chapter.

7. Confirm the request with the

OK

softkey.

The adjustment program starts and heats the heated lid to 105 °C. You will then be

prompted to open the lid.

8. Open the heated lid.
9. Confirm the request with the

OK

softkey.

You will be prompted to insert the temperature sensor (

Insert sensor in the position

indicated

).

10. Insert the temperature sensor in the position indicated.
11. Confirm the request with the

OK

softkey.

You will be prompted to close the heated lid.

12. Close the heated lid.
13. Confirm the request with the

OK

softkey.

The measuring cycle begins by tempering the thermoblock to 95 °C and 35 °C.