beautypg.com

3 safety, 1 intended use, 2 user profile – Eppendorf Temperature Verification System User Manual

Page 10: 3 information on product liability, 4 warnings for intended use

background image

Temperature Verification System USB – Single-channel — Operating manual

10

3

Safety

3 Safety

3.1 Intended use

The Temperature Verification System USB – Single-channel is a single-channel temperature

measuring device for verification and adjustment of thermoblock temperature control and

recording the temperature data of thermal cyclers. It supports the verification and adjustment of

the following Eppendorf product families:

Mastercycler nexus

Mastercycler pro

Mastercycler ep

Thermomixer

ThermoStat

The Mastercycler gradient and Mastercycler ep realplex can no longer be verified and adjusted

with the Temperature Verification System USB – Single-channel.
The Temperature Verification System USB – Single-channel may only be used indoors.

3.2 User profile

The device and accessories may only be operated by trained and skilled personnel.
Before using the device, read the operating manual carefully and familiarize yourself with the

device's mode of operation.

3.3 Information on product liability

In the following cases, the designated protection of the device may be compromised. Liability for

any resulting property damage or personal injury is then transferred to the operator:

The device is not used in accordance with the operating manual.

The device is used outside of its intended use.

The device is used with accessories or consumables which are not recommended by

Eppendorf.

The device is maintained or repaired by people not authorized by Eppendorf.

The user makes unauthorized changes to the device.

3.4 Warnings for intended use

Read the operating manual and observe the following safety instructions before using the

Temperature Verification System USB – Single-channel.

DANGER!

DANGER! Risk of explosion.

Do not operate the device in areas where work is completed with

explosive substances.

WARNING!

WARNING! Risk from incorrect supply voltage

Only connect the device to voltage sources which correspond to the

electrical requirements on the name plate.