5.1.3

Verifying the Mastercycler ep

Prerequisites

The Mastercycler is switched on.

The Measuring device is connected to the temperature sensor and switched on.

1. Log in as

User

or

Administrator

.

2. Select

System

menu item under the cycler nodes.

3. Press the

Verification

softkey.

A note on the Temperature Verification System USB – Single-channel will be shown.

4. Confirm the note with the

OK

softkey.

You will be asked if you would like to use the Temperature Verification System USB –

Single-channel.

5. Answer the question using the

Yes

softkey.

You will be prompted to connect the Temperature Verification System USB – Single-channel.

6. Connect and switch on the Temperature Verification System USB – Single-channel.

The connection of the Temperature Verification System USB – Single-channel to the various

Mastercycler devices is described in the installation chapter.

7. Confirm the request with the

OK

softkey.

The verification program starts and heats the heated lid to 105 °C. You will then be prompted

to open the lid.

8. Open the heated lid.
9. Confirm the request with the

OK

softkey.

You will be prompted to insert the temperature sensor (

Insert sensor in the position

indicated

).

10. Insert the temperature sensor in the position indicated.
11. Confirm the request with the

OK

softkey.

You will be prompted to close the heated lid.

12. Close the heated lid.

Hint!

Any Mastercycler ep connected to a control panel must be in the "Idle" state.

No programs may be running.