Positioning the silicone mat

Place the silicon mat over the spacers and the sensor so the thermoblock is evenly covered.

Close the heated lid.

4.6.4

Removing oil

Remove the oil after the last measurement.
The procedure is described in the cleaning chapter (see Removing oil from the thermoblock

on p. 33).

4.7 Connecting the Mastercycler to a computer and CycleManager pro/pro

XL

1. Use the temperature sensor connecting cable to connect the grip to the Measuring device.
2. Use the USB cable to connect the Measuring device and computer.
3. Switch on the Measuring device.
4. Check the rechargeable battery capacity.
5. Charge the rechargeable battery or use the mains/power supply device if the battery capacity

is low.

4.8 Installing the Thermomixer and ThermoStat

4.8.1

Preparing the Thermomixer comfort and ThermoStat plus

1. Mount the 1.5 mL thermoblock on the device to be verified if another thermoblock is mounted

on the device.

2. Switch on the device.

4.8.2

Connecting the Thermomixer and ThermoStat

1. Use the temperature sensor connecting cable to connect the grip to the Measuring device.
2. Switch on the Measuring device.
3. Check the rechargeable battery capacity.
4. Charge the rechargeable battery or use the mains/power supply device if the battery capacity

is low.