Temperature Verification System USB – Single-channel — Operating manual

23

13. Confirm the request with the

OK

softkey.

The measuring cycle begins by tempering the thermoblock to 95 °C and 35 °C.

14. After each temperature control, read the temperature value of the measuring device and

enter it in the control panel twice.
Each additional measuring cycle begins with the request to insert the temperature sensor in

a different bore. The position will be shown on the display.

15. Repeat steps 8 to 14.

The protocol will be shown on the display after verification.
The

StoreMMC

softkey is used to save the protocol.

The

Print

softkey is used to print the protocol.

16. Press the

Exit

softkey.

The

Cycler System Level

window will be shown.

5.1.4

Verifying the Mastercycler pro/ep using the CycleManager pro

Prerequisites

The Mastercycler is switched on.

The Measuring device with the temperature sensor is connected to the computer and

switched on.

1. Start the CycleManager pro/pro

XL

.

2. Log in as

User

or

Administrator

.

Hint!

All other functions of the CycleManager pro/pro

XL

are deactivated during the

verification.

Hint!

The current verification can be interrupted at any time using the icon.