Eppendorf Temperature Verification System User Manual
Page 26

Temperature Verification System USB – Single-channel — Operating manual
26
1. Log in as
Administrator
.
2. Select the
System
menu item under the cycler node.
3. Press the
Adjustment
softkey.
The adjustment program starts and heats the heated lid to 105 °C. You will then be
prompted to insert the temperature sensor (
Insert sensor in the position indicated
).
4. Open the heated lid.
5. Insert the temperature sensor in the bore at the displayed position.
6. Close the heated lid.
7. Start the measuring cycle using the Enter key.
The measuring procedure runs at 95 °C and then at 35 °C.
Each additional measuring cycle begins with the request to insert the temperature sensor in
a different bore. The position will be shown on the display.
8. Repeat steps 3 to 5.
9. Start each new measuring cycle using the OK softkey.
The protocol will be shown on the display after the adjustment.
10. Check the adjustment via a subsequent verification (see Verifying the Mastercycler nexus
and Mastercycler pro on p. 20).
11. Press the
Exit
softkey.
The
Cycler System Level
window will be shown.