3 preparing the thermoblock 384 – Eppendorf Temperature Verification System User Manual

Page 17

background image

Temperature Verification System USB – Single-channel — Operating manual

17

4.6.3

Preparing the Thermoblock 384

Prerequisites

The thermoblock with 384 wells is cooled to room temperature.

Fill the measuring positions with oil.

Inserting the spacers

Inserting the temperature sensor

Insert the temperature sensor after the program started.
The position and the request to insert the sensor will be shown in the display.
The sensor cable will be guided outward over the thermoblock on the front of the

Mastercycler. It should be freely movable and should not be positioned near the lid closing

mechanism.

Hint!

Oil in the space between the temperature sensor and the thermoblock

improves the heat transmission. For measurements with the Temperature

Verification System USB – Single-channel, the bores are filled with 5 μL oil at

the specified measuring positions.

Fill the bores with 5 μL oil at

the following positions:

F19

N22

E11

J14

C3

K6

Insert the spacers before starting the

program.
The spacers will be inserted at the following

positions:

B5

B20

O5

O20

A

B

C

D

E

F

G

H

I

J

K

L

M

N

O

P

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24