Eppendorf Temperature Verification System User Manual

Page 29

background image

Temperature Verification System USB – Single-channel — Operating manual

29

3. In the navigation tree, open the node of the cycler to be adjusted.
4. Select the

System

menu item.

5. Select the

Adjustment

tab.

6. Start the adjustment using the icon.
7. Select the Temperature Verification System USB – Single-channel using the checkbox.
8. Use the icon to confirm the selection.

You will be prompted to connect the Temperature Verification System USB – Single-channel.

9. Connect the Temperature Verification System USB – Single-channel.
10. Switch on the Measuring device.

The temperature sensor will be checked by the Mastercycler.
If the certification of the temperature sensor is no longer valid, an error message will be

shown. End the adjustment. Arrange to have the temperature sensor certified.

11. Confirm the query on the screen using the icon.

The adjustment program starts and heats the heated lid to 105 °C. You will then be

prompted to insert the temperature sensor (

Insert sensor in the position indicated

).

12. Open the heated lid.

13. Insert the temperature sensor in the position indicated.
14. Close the heated lid.

If a motor-driven heated lid is mounted, it automatically moves to the front position during the

next step.

15. Start the first measuring cycle using the icon.

The measuring procedure runs at 95 °C and then at 35 °C.Each additional measuring cycle

begins with the request to move the temperature sensor to a different position.

16. Repeat steps 13 to 16.
17. Start each new measuring cycle using the icon.

The protocol will be shown on the monitor after the adjustment.

18. Complete the adjustment procedure using the icon.