beautypg.com

3 disinfection/decontamination, 4 service and maintenance, Disinfection/decontamination – Eppendorf epMotion 96 User Manual

Page 66: Service and maintenance

background image

Maintenance
epMotion

®

96

English (EN)

66

8.3

Disinfection/Decontamination

1. Switch off the device and disconnect it from the mains/power line.

2. Remove all cables and accessories from the device.

3. Clean the device with a mild cleaning agent before disinfection.

4. Choose a disinfection method that corresponds to the legal regulations and guidelines in effect for your

range of application.

5. Wipe the surfaces with a lintfree cloth which is moisturized with a disinfectant.

6. Disinfect all parts that will be sent in with the device.

8.4

Service and maintenance

The Eppendorf AG service team is available to service and certify your device.

Eppendorf AG recommends annual servicing of the device.

Service provisions:

• Service
• Operational qualification (OQ) according to manufacturer's specifications
• Electrical safety testing according to country-specific regulations
• Software update

Information on the services offered can be found on our webpage: www.eppendorf.com/epservices.

DANGER! Electric shock as a result of penetration of liquid.

 Switch off the device and disconnect the power plug before starting cleaning or

disinfection work.

 Do not allow any liquids to penetrate the inside of the housing.
 Do not spray clean/spray disinfect the housing.
 Only plug the device back in if it is completely dry, both inside and outside.

For example, use alcohol (ethanol, isopropanol) or alcohol-based disinfectants.