beautypg.com

5 changing the liquid type, Changing the liquid type – Eppendorf epMotion 96 User Manual

Page 58

background image

Operation
epMotion

®

96

English (EN)

58

6.15.5

Changing the liquid type

Liquid properties are defined for individual pipette tip sizes. Calibration values may be entered for a
maximum of 5 different volume ranges. The higher the calibration value, the greater the piston stroke per
microliter of the set pipetting volume.

The calibration values of the liquids must be determined empirically. This can be done using a
gravimetric comparison measurement, for example.

It is not possible to change the

Standard

liquid type.

1. Select a liquid type and confirm with

Select

.

The system opens the

Liquid Type Editor

.