7 troubleshooting, 1 general errors, 1 pipette tips – Eppendorf epMotion 96 User Manual
Page 63: 2 wireless connection, Troubleshooting 7.1, General errors 7.1.1, Pipette tips, Wireless connection, 7troubleshooting 7.1 general errors
63
Troubleshooting
epMotion
®
96
English (EN)
7
Troubleshooting
7.1
General errors
7.1.1
Pipette tips
7.1.2
Wireless connection
Problem
Cause
Solution
Liquid is dripping from
the pipette tips.
• Wrong pipette tips.
Check the pipette tips.
Insert epT.I.P.S Motion Reloads 50
μL or 300 μL.
• Pipette tips do not fit the tip cones
properly.
Unlock the pipette tips.
Check the interspace.
Remove any foreign matter.
Lock the pipette tips.
• O-rings are damaged.
Contact the authorized service.
Pipette tips are not
recognized.
• Tray with pipette tips has been
rotated by 180°.
Insert the tray with the notch
pointing towards the handle of the
loading frame.
Problem
Cause
Solution
Connection error
• The device is switched off.
Switch the device on.
Shut down and restart the software.
• No wireless connection.
Open
Settings
on the iPod.
Activate
Wi-Fi
.
Shut down and restart the software.
• Wireless connection is not
configured correctly.
Open
Settings
on the iPod.
Open
Wi-Fi
and verify the network
settings.
Shut down and restart the software.
• No wireless connection.
• The distance between the iPod and
the device is too large.
Place the iPod closer to the device
or put it into the docking station.