2 moving the lifting table into basic position, 3 setting the stop for the lifting table, 5 aspirating liquid – Eppendorf epMotion 96 User Manual
Page 35: Moving the lifting table into basic position, Setting the stop for the lifting table, Aspirating liquid
35
Operation
epMotion
®
96
English (EN)
6.4.2
Moving the lifting table into basic position
6.4.3
Setting the stop for the lifting table
The stop defines the end position of the lifting table in the working position. It makes sense to use the stop
when filling several plates.
Prerequisites
• The stop is in the basic position.
6.5
Aspirating liquid
Prerequisites
• The device is switched on.
• The software has been started.
• The pipette tips are attached.
1. Put the source vessel onto the lifting table.
2. Select the operating mode.
3. Set the aspiration volume.
4. Set the parameters.
5. Move the lifting table into the working position until the pipette tips are immersed in the liquid.
1. Hold the lifting table and loosen the locking
handle.
2. Move the lifting table into the basic position.
Source vessel or destination vessel can be
replaced.
Pipette tips can be inserted.
1. Move the lifting table into working position.
2. Tighten the locking handle on the lifting table.
3. Tighten the locking handle for the stop.
The working position is fixed.