3 liquid volume, 4 file names, 5 sensors – Eppendorf epMotion 96 User Manual
Page 64: Liquid volume, File names, Sensors
Troubleshooting
epMotion
®
96
English (EN)
64
7.1.3
Liquid volume
7.1.4
File names
7.1.5
Sensors
Problem
Cause
Solution
Pipetting volume is
larger than ...
• Set liquid volume is too high.
Reduce the liquid volume.
• Pipette tips are too small.
Insert pipette tips with 50μL or
300μL capacity.
No tips installed
• No pipette tips inserted.
Insert epT.I.P.S Motion Reloads 50
μL or 300 μL.
• Pipette tips not recognized.
• Pipette tips were not properly
locked.
Push the lever all the way back.
Problem
Cause
Solution
File name to long
• File name is too long.
• Program cannot be saved.
Choose a file name with a
maximum of 21 characters.
File name already exists
• File name is already used.
• Program cannot be saved.
Choose a different file name.
Problem
Cause
Solution
Lower sensor triggered
• Sensor for the piston movement
was triggered.
Switch the device off and back on
again.
If the error persists, please contact
the authorized service.
Upper sensor triggered
• Device error.
• Sensor for the piston movement
was triggered.
Switch the device off and back on
again.
If the error persists, please contact
the authorized service.