Page 64: Liquid volume, File names, Sensors

background image

Troubleshooting
epMotion

®

96

English (EN)

64

7.1.3

Liquid volume

7.1.4

File names

7.1.5

Sensors

Problem

Cause

Solution

Pipetting volume is
larger than ...

• Set liquid volume is too high.

 Reduce the liquid volume.

• Pipette tips are too small.

 Insert pipette tips with 50μL or

300μL capacity.

No tips installed

• No pipette tips inserted.

 Insert epT.I.P.S Motion Reloads 50

μL or 300 μL.

• Pipette tips not recognized.
• Pipette tips were not properly

locked.

 Push the lever all the way back.

Problem

Cause

Solution

File name to long

• File name is too long.
• Program cannot be saved.

 Choose a file name with a

maximum of 21 characters.

File name already exists

• File name is already used.
• Program cannot be saved.

 Choose a different file name.

Problem

Cause

Solution

Lower sensor triggered

• Sensor for the piston movement

was triggered.

 Switch the device off and back on

again.

 If the error persists, please contact

the authorized service.

Upper sensor triggered

• Device error.
• Sensor for the piston movement

was triggered.

 Switch the device off and back on

again.

 If the error persists, please contact

the authorized service.