background image

33

Operation

epMotion

®

96

English (EN)

6.2.4

Exiting the simulation mode

6.3

Inserting pipette tips

1. Tap twice on the

Home

button of the iPods.

2. Swipe the software screen upwards out of the

window.
The simulation mode is exited.
The software is exited.

NOTICE! Dispensing error due to incorrect handling of pipette tips.
Tips become deformed and change size during autoclaving.

 Do not autoclave the pipette tips. Use tips with the sterile specification, if required.
 Do not stack any racks that contain pipette tips.

NOTICE! Damage to the device due to incorrectly equipped trays.
The device detects the tip size using the coding of the trays. Incorrectly equipping the trays
can cause contamination of pipetting head.

 Only use trays that were equipped by the manufacturer.
 Do not equip the trays manually.

When ejecting the pipette tips, liquid can drip from the pipette tips.
Make sure that the pipette tips are completely empty before replacing them.