beautypg.com

3 inserting a 384-well plate, 4 filling the 384-well plate, Inserting a 384-well plate – Eppendorf epMotion 96 User Manual

Page 38: Filling the 384-well plate

background image

Operation
epMotion

®

96

English (EN)

38

6.7.3

Inserting a 384-well plate

Prerequisites

• The lifting table is in the basic position.

6.7.4

Filling the 384-well plate

In order to fill a 384-well plate completely, the plate must be positioned consecutively in each corner of the
lifting table.

1. Lift the lifting table and take the adapter out of

the base.

2. Insert the adapter into the lifting table.
3. Insert the 384-well plate in the lower right-hand

corner.
Fill the plate beginning with well A1.

1. Place the 384-well plate in the bottom right

position.

2. Move the lifting table into working position.

Every second row and every second well in the
series starting at A1 is filled.

3. Move the lifting table into basic position.
4. Place the 384-well plate in the bottom left

position.

5. Move the lifting table into working position.

Every second row and every second well in the
series starting at A2 is filled.

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P

1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24

A
B
C
D
E
F
G
H
I
J
K
L
M
N
O