3 inserting a 384-well plate, 4 filling the 384-well plate, Inserting a 384-well plate – Eppendorf epMotion 96 User Manual
Page 38: Filling the 384-well plate
Operation
epMotion
®
96
English (EN)
38
6.7.3
Inserting a 384-well plate
Prerequisites
• The lifting table is in the basic position.
6.7.4
Filling the 384-well plate
In order to fill a 384-well plate completely, the plate must be positioned consecutively in each corner of the
lifting table.
1. Lift the lifting table and take the adapter out of
the base.
2. Insert the adapter into the lifting table.
3. Insert the 384-well plate in the lower right-hand
corner.
Fill the plate beginning with well A1.
1. Place the 384-well plate in the bottom right
position.
2. Move the lifting table into working position.
Every second row and every second well in the
series starting at A1 is filled.
3. Move the lifting table into basic position.
4. Place the 384-well plate in the bottom left
position.
5. Move the lifting table into working position.
Every second row and every second well in the
series starting at A2 is filled.
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P