beautypg.com

2 dispensing liquid, 3 semiautomatically dispensing liquid, Dispensing liquid – Eppendorf epMotion 96 User Manual

Page 42: Semiautomatically dispensing liquid

background image

Operation
epMotion

®

96

English (EN)

42

6.9.2

Dispensing liquid

Prerequisites

• Liquid has been aspirated.
• Destination vessels are available.

1. Put the destination vessel onto the lifting table and move the lifting table into working position.

2. Press

Dispense

.

3. Change or reposition the destination vessel.

6.9.3

Semiautomatically dispensing liquid

It is advisable to use the

Auto

function when filling a single 384-well plate or several 96-well plates in

succession.

1. Press the

Auto

button.

2. Set the dispensing volume.

3. Set the number of steps for dispensing.

4. Aspirate liquid from a source vessel

5. Put the destination vessel onto the lifting table.

6. Move the lifting table into working position and tighten the locking handle.

7. Tighten the locking handle for the stop.

8. Move the lifting table into basic position.

9. Change or reposition the destination vessel.

10. Move the lifting table against the stop and keep it in this position until the dispensing of the liquid is

initiated.

The number of the remaining dispensing steps is displayed.