Product description
epMotion

®

96

English (EN)

14

2.4

Recommended pipette tips

We recommend using epT.I.P.S Motion Reloads. The pipette tips are available with various degrees of
purity, with and without filters.

2.5

Recommended plates

NOTICE! Dispensing error due to incorrect handling of pipette tips.
Tips become deformed and change size during autoclaving.

 Do not autoclave the pipette tips. Use tips with the sterile specification, if required.
 Do not stack any racks that contain pipette tips.

Pipette tip

Sizes

Use

Plate depth for liquid
aspiration

epT.I.P.S Motion
Filter Reloads

50 μL

Liquids that require the pipette to
be protected from contamination by
aerosols.

max. 37 mm

300 μL

max. 34 mm

epT.I.P.S Motion
Reloads

50 μL

Liquids without any special risks.

max. 37 mm

300 μL

max. 34 mm

Plate

Size Use

Microplate

96 or 384 wells

Supplying and aspirating liquids (e.g. assays for plate
reader)

Deepwell plate

96 or 384 wells

Supplying and aspirating liquids (e.g. for culturing
organisms or cell cultures)

PCR plate

96 or 384 wells

Supplying and aspirating liquids (e.g. for PCR applications)

Reservoir

1, 8, or 12 chambers

Supplying liquids