6.15.2
Set the pipetting properties
Abb. 6-10: Pipetting Settings
Fig. 6-10:
Pipetting Settings
6.15.3
Liquid Types
– Setting the liquid properties
The liquid properties allow adjusting the pipetting to liquids with densities or vapour pressures that differ
sigificantly from those of water. The preset
Standard
is used for verifying the device and is suitable for
water and diluted aqueous solutions.
Parameter
Value
Pipette tips
Default settings
Reverse stroke (μL)
Setting the additional volume for the
liquid aspiration in μL.
50 μL
3 μL
300 μL
12 μL
Remaining stroke (μL)
Setting the residual volume in μL for
liquid aspiration in
Multidispense
mode.
50 μL
4 μL
300 μL
12 μL
Blow-out (uL)
Setting the air volume in μL for
dispensing the residual liquid.
50 μL
15 μL
300 μL
20 μL
Piston Pause (s)
Setting the delay time between pipetting
operations in seconds.
–
1 s
Auto Delay (s)
Setting the delay time for automatic
liquid dispensing in seconds
(
Multidispense
mode).
–
1 s