beautypg.com

2 set the pipetting properties, 3 liquid types – setting the liquid properties, Set the pipetting properties – Eppendorf epMotion 96 User Manual

Page 56: Liquid types – setting the liquid properties

background image

Operation
epMotion

®

96

English (EN)

56

6.15.2

Set the pipetting properties

Abb. 6-10: Pipetting Settings

Fig. 6-10:

Pipetting Settings

6.15.3

Liquid Types

– Setting the liquid properties

The liquid properties allow adjusting the pipetting to liquids with densities or vapour pressures that differ
sigificantly from those of water. The preset

Standard

is used for verifying the device and is suitable for

water and diluted aqueous solutions.

Parameter

Value

Pipette tips

Default settings

Reverse stroke (μL)

Setting the additional volume for the
liquid aspiration in μL.

50 μL

3 μL

300 μL

12 μL

Remaining stroke (μL)

Setting the residual volume in μL for
liquid aspiration in

Multidispense

mode.

50 μL

4 μL

300 μL

12 μL

Blow-out (uL)

Setting the air volume in μL for
dispensing the residual liquid.

50 μL

15 μL

300 μL

20 μL

Piston Pause (s)

Setting the delay time between pipetting
operations in seconds.

1 s

Auto Delay (s)

Setting the delay time for automatic
liquid dispensing in seconds
(

Multidispense

mode).

1 s