beautypg.com

BNC 835-3 Programmer Manual User Manual

Page 31

background image

Berkeley Nucleonics
www.Berkeley Nucleonics.com

835-3

Programmer’s Manual v1.0

Page 31 of 37

GATE : trigger level starts/stops sweep

2. Setting the waveform’s response to triggers:

— CONTinuous : reapeatedly accepts trigger events
— SINGle : uses only one trigger event


3. Selecting the trigger source which determines how the BNC MODEL 835-3 receives its
trigger signal, internally or externally. The GATE choice requires an external trigger.

3. Setting the trigger polarity when using an external source

Command

Parameters

Unit
(default)

[SOURce]:TRIGger[:SEQuence]:TYPE

NORMal | GATE | POINT

[SOURce]:TRIGger[:SEQuence]:TYPE:GATE

LOW|HIGH

HIGH

[SOURce]:TRIGger[:SEQuence]:SOURce

IMMediate|KEY|EXT|BUS

IMM

[SOURce]:TRIGger[:SEQuence]:DELay

S

[SOURce]:TRIGger[SEQuence]:SLOPe

POSitive|NEGative

POS

[SOURce]:RETRigger

ON|OFF|IMMediate

Not

impleme

nted

[SOURce]:TRIGger[:SEQuence]:ECOunt

1


:TRIGger:TYPE
[SOURce]:TRIGger:TYPE NORMal | GATE | POINT
[SOURce]:TRIGger:TYPE?
This command sets the trigger type that controls the waveform’s playback.

The following list describes the trigger type command choices:
NORMal

Upon triggering, the waveform sequence plays according to
settlings defined by :INITiate:CONTinuous (only once or repeatedly)

GATE

An external trigger signal repeatedly starts and stops the
waveform’s playback. The time duration for playback depends on
the duty period of the trigger signal and the gate polarity selection.
The waveform plays during the inactive state and stops during the
active polarity selection state. The active state can be set high or
low. The gate mode works only with an external trigger source.

POINT

Upon triggering, only a single point of the sweep (list) is played.


*RST NORM


:TRIGger:TYPE:GATE
[SOURce]:TRIGger:TYPE:GATE LOW|HIGH
[SOURce]:TRIGger:TYPE:GATE?
This command selects the active state (gate polarity) of the gate while using the gating
trigger mode. The LOW and HIGH selections correspond to the low and high states of an
external trigger signal. For example, when you select HIGH, the active state occurs
during the high of the trigger signal.
When the active state occurs, the BNC MODEL 835-3 starts the waveform playback at the
last played sample point, then stops the playback at the next sample point when the
inactive state occurs.
LOW

The waveform playback starts when the trigger signal goes low (active
state) and stops when the trigger signal goes high (inactive state).