BNC 835-3 Programmer Manual User Manual
Page 19
![background image](/manuals/545976/19/background.png)
Berkeley Nucleonics
www.Berkeley Nucleonics.com
835-3
Programmer’s Manual v1.0
Page 19 of 37
Continuously rearms the BNC MODEL 835-3 trigger system after completion of a
triggered
sweep.
:OUTPut Subsystem
Command
Parameters
Unit
(default)
Remark
OUTPut[:STATe]
ON | OFF | 1 | 0
OFF
OUTPut:BLANking
ON | OFF | 1 | 0
OFF
FW2.4
[:STATe] ON|OFF|1|0
:OUTPut[:STATe] ON|OFF|1|0
Turns BNC MODEL 835-3 RF output power on/off.
:BLANking[:STATe] ON|OFF|1|0
:OUTPut:BLANking[:STATe] ON|OFF|1|0
:OUTPut:BLANking[:STATe]?
ON causes the RF output to be turned off (blanked) during frequency changes. OFF
leaves RF output turned on (unblanked).
[SOURce]:FREQuency Subsystem
Command
Parameters
Unit
(default)
[SOURce]:FREQuency:FIXed | CW
fmin to fmax
Hz
[SOURce]:FREQuency:MODE
FIX | CW | SWEep | LIST
FIXed
[SOURce]:FREQuency:STARt
fmin to fmax
Hz
[SOURce]:FREQuency:STOP
fmin to fmax
Hz
[SOURce]:FREQuency:STEP[:LINear]
0 to fmax-fmin
Hz
[SOURce]:FREQuency:STEP:LOGarithmic
0 to 255.999999999
:FREQuency[:CW]
[SOURce]:FREQuency[:CW]
[SOURce]:FREQuency[:CW]?
This command sets the signal generator output frequency for the CW frequency mode.
*RST 100 MHz
Range Please refer to the Data Sheet.
:FREQuency:MODE
[SOURce]:FREQuency:MODE FIX|CW|SWEep|LIST
[SOURce]:FREQuency:MODE?
This command sets the frequency mode of the signal generator to CW or swept.
FIX | CW
Stops a frequency sweep
SWEep or LIST
This choice selects the swept frequency mode. If sweep triggering is
set to immediate along with continuous sweep mode, executing the
command starts the LIST or SWEep frequency sweep. In SWEep
mode, frequency will be determined by programmed values for the
:STARt and :STOP :FREQuency subsystem commands. In LIST