beautypg.com

BNC 835-3 Programmer Manual User Manual

Page 19

background image

Berkeley Nucleonics
www.Berkeley Nucleonics.com

835-3

Programmer’s Manual v1.0

Page 19 of 37

Continuously rearms the BNC MODEL 835-3 trigger system after completion of a
triggered
sweep.

:OUTPut Subsystem

Command

Parameters

Unit

(default)

Remark

OUTPut[:STATe]

ON | OFF | 1 | 0

OFF

OUTPut:BLANking

ON | OFF | 1 | 0

OFF

FW2.4


[:STATe] ON|OFF|1|0
:OUTPut[:STATe] ON|OFF|1|0
Turns BNC MODEL 835-3 RF output power on/off.


:BLANking[:STATe] ON|OFF|1|0
:OUTPut:BLANking[:STATe] ON|OFF|1|0
:OUTPut:BLANking[:STATe]?
ON causes the RF output to be turned off (blanked) during frequency changes. OFF
leaves RF output turned on (unblanked).

[SOURce]:FREQuency Subsystem

Command

Parameters

Unit

(default)

[SOURce]:FREQuency:FIXed | CW

fmin to fmax

Hz

[SOURce]:FREQuency:MODE

FIX | CW | SWEep | LIST

FIXed

[SOURce]:FREQuency:STARt

fmin to fmax

Hz

[SOURce]:FREQuency:STOP

fmin to fmax

Hz

[SOURce]:FREQuency:STEP[:LINear]

0 to fmax-fmin

Hz

[SOURce]:FREQuency:STEP:LOGarithmic

0 to 255.999999999


:FREQuency[:CW]
[SOURce]:FREQuency[:CW]
[SOURce]:FREQuency[:CW]?
This command sets the signal generator output frequency for the CW frequency mode.
*RST 100 MHz
Range Please refer to the Data Sheet.

:FREQuency:MODE
[SOURce]:FREQuency:MODE FIX|CW|SWEep|LIST
[SOURce]:FREQuency:MODE?
This command sets the frequency mode of the signal generator to CW or swept.
FIX | CW

Stops a frequency sweep

SWEep or LIST

This choice selects the swept frequency mode. If sweep triggering is
set to immediate along with continuous sweep mode, executing the
command starts the LIST or SWEep frequency sweep. In SWEep
mode, frequency will be determined by programmed values for the
:STARt and :STOP :FREQuency subsystem commands. In LIST