This command sets the power range extension attenuator. This command will also turn
off automatic attenuation setting. Refer to
[SOURce]:POWer:ATTenuation:AUTO OFF
for
details. Attenuator input RF power shall be -30…+10 dBm, e.g. using a
[SOURce]:POWer:ATTenuation value of 50 dB, the output RF power range is -80…-40
dBm.
:ATTenuation:AUTO
[SOURce]:POWer:ATTenuation:AUTO ON|OFF|1|0
[SOURce]:POWer:ATTenuation:AUTO?
(Devices with option PE only)
This command turns the power range extension on or off. Turning it off allows fast power
sweeps for devices featuring an extended output power range.
:ATTenuation:LIST?
[SOURce]:POWer:ATTenuation:LIST?
(Devices with option PE only)
This query returns a comma-separated list of available attenuation settings. These can be
set using the
[SOURce]:POWer:ATTenuation
command.
Available settings are:
0, 35, 70 dB for option PE
0, 10, 20, …, 110 dB for option PE2
[SOURce]:ROSCillator Subsystem
Command
Parameters
Unit
(default)
[SOURce]:ROSCillator:EXTernal:FREQuency
100 MHz
[SOURce]:ROSCillator:SOURce
INTernal|EX
Ternal
INT
:REFerence:EXTernal:FREQuency
[SOURce]:ROSCillator:EXTernal:FREQuency
This command conveys the expected reference frequency value of an externally applied
reference the signal generator. The variable
*RST 10 MHz
Range 1 to 100 MHz in 1 MHz steps
: SOURce
[SOURce]:ROSCillator: SOURce INT|EXT
This command selects either an internal or external reference for the waveform clock.
*RST INT
[SOURce]:LIST Subsystem
Command
Parameters
Unit
(default)
Remark
[SOURce]:LIST:DIRection
UP|DOWN|RANDom
UP
[SOURce]:LIST:DWELl
s
[SOURce]:LIST:DELay
s
[SOURce]:LIST:DELay:AUTO
ON|OFF|1|0
ON
[SOURce]:LIST:FREQuency:POINts?
[SOURce]:LIST:FREQuency
Hz