Stat – BNC 745-20C-T User Manual
Page 24

User's Manual - MOD745-20
October 2014
Page 24 / 30
6.2.10. STAT
Syntax:
STAT CLEAR
STAT?
Form:
Set & Query
Description:
Query equipment information
Parameter:
CLK (Clock source, internal or external)
POW (+6V, -6V, +3.3V, +1.8V, +11V power supply level in V)
TRIG (Trigger feedback for channel 1 to 20. 1 = trigged)
MTRIG (Single shot trigger feedback. 1 = trigged)
RST value:
-
Example:
Clear information: STAT CLEAR
Query temperature: STAT? TEMP (Answer: :STAT TRET,28.30)
Query clock source: STAT? CLK (Answer: :STAT CLK,INTERNAL)
Query power supply level: STAT? POW
(Answer: :STAT POW,6.00,-5.99,3.35,1.75,11.45)
Query trigger feedback: STAT? TRIG
(Answer: :STAT TRIG,1,0,0,1,0,0,0,0,0,0,0,0,0,0,0,1,1,0,0,0)
Query single shot trigger feedback: STAT? MTRIG
(Answer: :STAT MTRIG,0)