6.2.3. TRIGGER

Syntax:

TRIG T,

TRIG? T

Form:

Set & Query

Description:

Query trigger mode on channel or set channel trigger mode to

Frequency 1 (F1), Frequency 2 (F2), Frequency 3 (F3), Single Shot 1

(SS1), Single Shot 2 (SS2) or Inhibit (INH)

Parameter:

: Channel number (0 to 20)

: Trigger mode (F1, F2, F3, SS1, SS2 or INH)

RST value:

INH

Example:

Set channel 2 trigger mode to Frequency 1: TRIG T2,F1

Query trigger mode on channel 2: TRIG? T2 (Answer: :TRIG T2,F1)

6.2.4. VOLTAGE LEVEL

Syntax:

AMPL T,

AMPL? T

Form:

Set & Query

Description:

Query voltage level on channel or set channel to voltage level

Parameter:

: Channel number (0 to 20)

: Voltage level (in millivolt)

RST value:

Last value set

Range:

3 000 to 6 000 mV

Example:

Set voltage level to 3.5 V on channel 4: AMPL T4,3500