6.2.5. WIDTH

Syntax:

WIDTH T,

WIDTH? T

Form:

Set & Query

Description:

Query channel width or set channel at specified width

Parameter:

: Channel number (0 to 20)

: Width (in nanosecond)

RST value:

Last value set

Range:

100 to 300 000 000 ns

Example:

Set 250 ns width on channel 4: WIDTH T4,250

Query width of channel 4: WIDTH? T4 (Answer: :WIDTH T4,2500)

6.2.6. POLARITY

Syntax:

TTL T,

TTL? T

Form:

Set & Query

Description:

Query polarity of channel or set channel at specified

polarity, positive (POS) or negative (NEG)

Parameter:

: Channel number (0 to 20)

: Polarity (POS or NEG)

RST value:

POS

Example:

Set negative-going pulses on channel 3: TTL T3,NEG