beautypg.com

Kramer Electronics VP-1201 User Manual

Page 32

background image

KRAMER: SIMPLE CREATIVE TECHNOLOGY

Connecting a VP-1201 12x1 XGA Switcher / Scanner

28

5.7 Setting the Dipswitches

This section describes the machine set-up and dipswitch selection.
By default, all the

VP-1201 dipswitches are set to OFF. Figure 27 and Table

8 describe the

VP-1201 unit dipswitches, which consist of two sets, each

numbered 1 to 8, respectively.

1 2

3 4 5 6

7

8

1 2

3 4 5 6

7

8

Figure 27: VP-1201 Dipswitches

Table 8: Dipswitch Settings

DIPS

Function

Description

1, 2, 3

MACH. ADDR.

Determines the position of the machine in the input expansion

configuration

4, 5, 6, 7 MACH. #

For allowing several cascaded systems to be interconnected for

control purposes

1

8

RS-485 TERM

ON for RS-485 Line Termination

OFF for no RS-485 Line Termination

1, 2, 3, 4 DWELL TIME

Determines scanning delay time – the number of seconds each

input is connected to the output in the Scanner mode

5

SCAN

ON for Scanner mode; OFF for switcher mode

6

VALID ONLY

ON determines automatic scanning of only valid signals

7

CASCADE

ON when in the Input Expansion configuration (see section 5.5.2)

8

FLASH PROG.

ON to enable upgrade to the latest Kramer firmware (see section 7)

1 A configuration may be built with several groups of cascaded machines. To do this, set the machine addresses accordingly
for each group, and set all the machines in a group with the same MACHINE #. For example, a system with a 24 input
switcher and 36 input switcher would consist of two switchers with machine # 1, and three switchers with machine # 2.
Machine # 1 would consist of a unit with MACHINE ADDRESS 1 and a unit with MACHINE ADDRESS 2. Machine # 2
would consist of units with MACHINE ADDRESSES 1, 2 and 3