Aalborg TIO Totalizer I/O Flow Monitor/Controller User Manual
Page 81
78
AALBORG
7
TOT
A
LIZER-IO ASCII SOFTW
ARE INTERF
ACE COMMANDS
N
OTE
:
A
N
"*"
INDICA
TES
POWER
U
P
D
EF
AUL
T
SETTINGS
.
A
N
"**"
INDICA
TES
OPTIONAL
FEA
TURE
NOT
A
V
A
ILABLE
O
N
A
LL
MODELS
.
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Flow
Requests the current flow
reading in current EU
1
F
alue> (Actual flow in current engineering units) Set Point Set point value in current EU (Read, Write) With Argument #2 = ‘S’ set point value will be saved in the EEPROM 2 S NO ARGUMENT (read status) S: alue> (Set Point in current engineering units) Example: S:20.5 alue> (write, not saved in EEP- ROM memor y) S: alue> (Set Point in current engineering units) Example: S:20.5 alue>(write, saved in EEPROM memor y) S S: alue> ,S (Set Point in current engineering units) Example: S:20.5,S Density Read / Set Fluid Density for standard conditions in g/litr [0.000001-10000.0] g/litr 3 D NO ARGUMENT (read current value) D: alue> (Actual density in g/litr) Example: D:1.25 alue> (write and save new value) D: alue> (Actual density in g/litr) Example: D:1.56 Diagnostic Events Register Read/Reset current status of Diagnostic Events Register See list of the Diagnostic Events below . 4 DE NO ARGUMENT (read status) DE:0x10 0x10 – diagnostic word (16 bits wide) R (reset Event Log register to 0x0000) DE:0x0