d) Parameters configuration
The following parameters must be configured:
• Device Function (See 5.3.12 Submenu “Device Function”).
“Controller” function must be selected.
• Full Scale Range (See 5.3.14 Submenu “Device Calibration”). Full
Scale Range parameter must be set equal to the GFC full scale flow
rate in Litr/min.
• Fluid Std. Density (See 5.3.14 Submenu “Device Calibration”). This
parameter is required only when mass-based engineering units are
selected.
User may configure other parameters (see Paragraph 5.3) according to his pref-
erences and application requirements.
6.2.3 Connecting TIO to flow meters/controllers
from other manufactures (stand alone)
a) Mounting
When TIO is connected to flow meters / controllers from other
manufactures, it can be used as stand-alone table top or panel-
mounted (see Figure 6.8). On the back side of the TIO enclosure there
are 4 tapped holes which are designated to be used for the panel-
mounted option.
NOTE: If “Full Scale Range”, “Device Function” and “Fluid Std.
Density” parameters are not set properly, the device may have
erroneous reading and unpredictable behavior.
,