5 module 5 - s, Serial ), 9600 baud – Red Lion CUB5T User Manual
Page 11: Bit data, Odd parity, 00 addr, No abbr, No prnt opt, Erial, Ommunications

11
DATA BIT
ABBREVIATED PRINTING
PRINT OPTIONS
PARITY BIT
METER ADDRESS
300
1200
600
2400
19200
38400
4800
9600
7-bit
8-bit
NO
Odd
EVEN
NO
YES
NO
YES
Module 5 is the programming module for the Serial Communications
Parameters. These parameters are used to match the serial settings of the
CUB5T with those of the host computer or other serial device. The Serial Setup
Parameters are only accessible when an optional RS232 or RS485 serial
communications module is installed in the meter.
This section replaces the bulletin shipped with the RS232 and RS485 serial
communications plug-in cards. Discard the separate bulletin when using those
serial plug-in cards with the CUB5T.
SEL
Data Bit
Baud Rate
Parity Bit
Options
Meter
Address
Abbreviated
Printing
5-SEriAL
bAUd
dAtA
PAritY
Addr
Abbr
Prnt OPt
Copy Program
Settings
COPY
Pro
6.5 MODULE 5 - S
ERIAL
C
OMMUNICATIONS
P
ARAMETERS
(5-SEriAL)
PARAMETER MENU
BAUD RATE
Set the baud rate to match that of other serial communications equipment.
Normally, the baud rate is set to the highest value that all of the serial
communications equipment is capable of transmitting and receiving.
Select either 7- or 8-bit data word length. Set the word length to match the
other serial communications equipment on the serial link.
This parameter only appears when the Data Bit parameter is set to a 7-bit
data word length. Set the parity bit to match that of the other serial equipment
on the serial link. The meter ignores parity when receiving data and sets the
parity bit for outgoing data. If parity is set to
NO
, an additional stop bit is used
to force the frame size to 10 bits.
Enter the serial node address. With a single unit, an address is not needed
and a value of zero can be used (RS232 applications). Otherwise, with multiple
bussed units, a unique address number must be assigned to each meter. The
node address applies specifically to RS485 applications.
This parameter determines the formatting of data transmitted from the meter
in response to a Transmit Value command or a Block Print Request. Select
NO
for a full print transmission, consisting of the meter address, mnemonics, and
parameter data. Select
YES
for abbreviated print transmissions, consisting of the
parameter data only. This setting is applied to all the parameters selected in the
PRINT OPTIONS. (Note: If the meter address is 0, the address will not be sent
during a full transmission.)
9600
bAUd
7-bit
dAtA
Odd
PAritY
00
Addr
NO
Abbr
NO
Prnt OPt
0
to
99
This parameter selects the meter values transmitted in response to a Print
Request. A print request is also referred to as a block print because more than
one parameter can be sent to a printer or computer as a block.
Selecting
YES
displays a sublist for choosing the meter parameters to appear
in the print block. All active parameters entered as
YES
in the sublist will be
transmitted during a block print. Parameters entered as
NO
will not be sent.
The “Print All” (
Prnt ALL
) option selects all meter values for transmitting
(
YES
), without having to individually select each parameter in the sublist.
Note: Inactive parameters will not be sent regardless of the print option
setting. For example, the Cycle Counter and Cycle Counter Start values will
only be sent when the Cycle Counter is enabled. If disabled, these parameters
are inactive and will not be transmitted. Likewise, the Setpoint parameters will
not be sent unless an optional setpoint card is installed in the meter.
DISPLAY
MNEMONIC
t-VALUE
Timer
YES
TMR
C-VALUE
Cycle Counter
NO
CNT
t-StOP
t-Strt
Timer Stop
Timer Start
NO
NO
TSP
TST
Cnt Strt
NO
CST
SPt ON
NO
SPT
SPt tOUt
SPt OFF
Setpoint Time-out
Setpoint OFF
NO
NO
STO
SOF
DESCRIPTION
FACTORY
SETTING
Counter Start
Setpoint ON