Static cfr calibration procedure (strip till) – Ag Leader InSight Strip-Till User Manual
Page 10

Part No. 2002831-42 Rev. C
10
Section 6: Strip Till Calibration
To perform the Static CFR Calibration, press the Static CFR Calibration button in the
Strip-Till Control Window.
Static CFR Calibration Procedure (Strip Till)
STEP
ACTION
Read Static Calibration warning
A warning appears, advising to disable the blower fan circuit and prepare to catch any
product dispensed in an appropriate container. Press OK.
1
Select Metering Circuit to Calibrate
The CFR Calibration Wizard appears. Choose the desired channel to calibrate, and
press Next.
2
Enter Dispense Amount
Use the numeric keypad to enter the amount of product to be dispensed into the
container, and press Next.
3
Enter Simulated Target Rate
Use the numeric keypad to enter a simulated target rate, shown in pounds per acres.
Press Next when finished.
4
Start Target Rate Countdown
Press the green Start button to begin the target rate countdown. As the countdown is
started, the button will turn red and state Stop. When the countdown is complete, press
the Next button.
5
Enter Actual Dispense Amount
Enter the actual dispense amount, in pounds.
Note: The CFR will be calculated from the actual product amount dispensed.
6
Calibration
Complete
A message appears, stating that your calibration is complete, and showing the CFR
amount, in cubic feet per revolution.
Either:
Press the Repeat Calibration button, or
Press Finish.