Prime User Manual r08

Page 17

5.4.2

Calibration Results

Once calibration is complete the “Calibration Results” window will indicate the quality

of the calibration. This applies to both magnetic sensor and accelerometer calibration.

The X, Y, and Z “Dist Coverage” values show a percentage of each vector that has been

covered during the calibration. The only way to get a Z value greater than 50% would be

to take some points with the module upside-down. The “Field Std Dev” values for Mag

Score and Accel Score indicate the overall quality of the calibration. The target value for

the “Field Std Dev” Mag Score is ~0.1 and the Accel Score value should be ~2, where a

lower score is better. Note that acceptable compass performance can be obtained with

somewhat higher scores, and the obtainable score will be a function of the host system,

the specific Prime module, and the execution of the calibration procedure. See Section

6.2 for additional information.

If a Mag Only Calibration is performed, the Mag Score will reflect the new magnetic

sensor calibration, while the Accel Score will be grayed out and represent the last saved

Accel Score values. Similarly, if an Accel Only Calibration is performed the Accel Score

will reflect the new accelerometer calibration and the Mag Score will be grayed out and

reflect the last saved Mag Score values.

If the button is clicked during a calibration process prior to taking the minimum

required samples, this will abort the calibration. (The button turns into the

button once the calibration process is started.) The scores associated with the

calibration method that was aborted will be “-1”, while scores unassociated with the

aborted calibration method will be grayed out and represent their last saved values.

If the calibration is acceptable, click the button to save the

calibration. If this button is not selected then the unit will need to be recalibrated after it

is turned off.

Note: The values

in μT or mg refer to the quality of the calibration and NOT the accuracy of the

heading. It is possible to have a good calibration but poor heading accuracy if the local magnetic
field changes after calibration.