beautypg.com

Calibration settings, 6 calibration settings – PNI CompassPoint Prime User Manual

Page 20

background image

PNI Sensor Corporation

DOC# 1014177 r08

Prime User Manual

– July 2011

Page 14

new data as soon as the previous request has been fulfilled. Note that the inverse of

the Poll Delay is somewhat greater than the sample rate, since the Poll Delay does not

include actual acquisition time.

Interval Delay

The Interval Delay is relevant when Push Mode is selected, and is the time delay, in

seconds, between completion of the Prime module sending one set of sampled data

and the start of sending the next sample set. If the time is set to 0 then the Prime will

begin sending new data as soon as the previous data set has been sent. Note that the

inverse of the Interval Delay is somewhat greater than the sample rate, since the

Interval Delay does not include actual acquisition time.

Acquire Delay

The Acquire Delay sets the time between samples taken by the module, in seconds.

This is an internal setting that is NOT tied to the time with which the module

transmits data to StudioPrime or the host system. Generally speaking, the Acquire

Delay is either set to 0, in which case the Prime is constantly sampling or set to equal

either the Poll Delay or Interval Delay values. The advantage of running with an

Acquire Delay of 0 is that the FIR filter can run with a relatively high Tap value to

provide stable and timely data. The advantage of using a greater Acquire Delay is

that power consumption can be reduced, assuming the Interval or Poll Delay are no

less than the Acquire Delay.

Flush Filters

Selecting this box results in the FIR filters being flushed (voided out) after each

measurement reading. This can be useful if a single, accurate reading is desired and

there is sufficient time to re-populate the filters The default is not to flush the filters.

Note: I

f “Flush Filters” is selected the rate at which the output is updated will be inversely

proportional to the number of taps. For example, if 8 taps and flush filters are selected, the
output will be updated at ~3.5 Hz, while it will be updated at ~1 Hz if 32 taps is selected.

5.3.6

Calibration Settings

The Prime supports both magnetic sensor and accelerometer calibration. These

calibrations can be performed independently or simultaneously. The relevant

StudioPrime inputs for these calibrations are discussed below. See Section 6.2 for

information on how to perform a calibration.

Stability Checking

By default when calibrating the module, a measurement must be stable for 3

consecutive readings prior to saving the sample for use in the calibration. This is why

the module must be held steady between points during the field calibration. This

stability helps to ensure a proper heading and allow for higher accuracy, but it also