Default, Revert, Enable 3d model – PNI CompassPoint Prime User Manual
Page 21
Prime User Manual r08
Page 15
takes more time. If this checkbox is deselected, then the module will NOT wait for a
stable reading and will immediately take a reading once the minimum change-
between-points threshold has been met.
Automatic Sampling
When this checkbox is selected, the module automatically takes a calibration point
once the minimum change-between-points requirement and the stability check
requirement (if selected) have been satisfied. If the user wants direct control over
when a calibration point is taken, then Automatic Sampling should be deselected. In
this case, the
the
minimum requirements are met.
Calibration Points
The user can select the number of points to take during a calibration. A minimum of
12 sample points are needed for a successful magnetic sensor calibration, while 18
samples are recommended for either accelerometer-only calibration or simultaneous
magnetic sensor and accelerometer calibration.
Mag Only Calibration
Select when only magnetic sensor calibration will be performed.
Accel Only Calibration
Select when only an accelerometer calibration will be performed.
Accel and Mag Calibration
Select when magnetic sensor and accelerometer calibrations will be performed
simultaneously.
5.3.7
Default
Clicking this button reverts the StudioPrime program to the factory default settings.
5.3.8
Revert
This button will have the StudioPrime program read the settings from the unit and display
them on the screen.
5.3.9
Enable 3D Model
StudioPrime’s Test tab includes a live-action 3-D rendering of a helicopter. Some
computer systems may not have the graphics capability to render the 3D Model, and for
this reason it may be desirable to turn off this feature.