FUTEK IPM500 (D500) Digital Display User Manual
Page 9

Press Menu
Select Key
Press Digit Select
Key
Press Value Select
Key
00000U
RS485 half or full duplex
0
Full duplex
1
Half duplex
0
0000U
Special start & stop char.
0
Standard continuous mode
1
Special start & stop characters
00
000U
RTS mode
0
Normal non-latching RTS
1
Single transmission, latching RTS
000
00U
Termination characters
0
Only at end of all items
1
At end of each item
0000
0U
Data sent, digital panel meter
only
0
Reading
1
Peak
2
Valley
3
Reading + Peak
4
Reading + Valley
5
Reading + Peak + Valley
.
SEr 3
.
Serial Setup 3
0000
0U
Data sent, scale meter only
0
Net + Gross
1
Net only
2
Gross only
3
Peak only (Net or Gross)
4
Net + Gross + Peak
5
Valley only
__
000U
Modbus ASCII gap timeout
0
1 sec
1
3 sec
2
5 sec
3
10 sec
__0
00U
Serial protocol
0
Custom ASCII
1
Modbus RTU
2
Modbus ASCII
SEr 4
.
Serial Setup 4
__00
0U
Parity
0
None
1
Odd
2
Even
_
Addr
Modbus Address
__
000 __000 __000
Select digit to flash.
__
158
Select
0
through
9
for flashing digit.
Address range is 1 to 247.
5.2 FRONT PANEL SETUP, COUNTERS & TIMERS ONLY
Press Menu
Select Key
Press Digit Select
Key
Press Value Select Key
___
000
Output filtering
0
Send unfiltered signal
1
Send filtered signal
___0
00U
Baud rate
0
300 baud
1
600 baud
2
1200 baud
3
2400 baud
4
4800 baud
5
9600 baud
6
19200 baud
.
Ser 1
.
Serial Setup 1.
Press
until
Ser 1
is displayed.
Fixed Parameters
- No parity
- 8 data bits
- 1 stop bit
___00
0U
Output update rate
60 Hz
50 Hz
0
0.017 sec
0.020 sec
1
0.28 sec
0.34 sec
2
0.57 sec
0.68 sec
3
1.1 sec
1.4 sec
4
2.3 sec
2.7 sec
5
4.5 sec
5.4 sec
6
9.1 sec
10.9 sec
7
18.1 sec
21.8 sec
8
36.6 sec
43.5 sec
9
72.5 sec
97 sec
__
0000U
Line feed
0
No
1
__0
000U
Alarm data with readings
0
No alarm data
1
Alarm data with reading
__00
00U
Control of data output
0
Continuous data output
1
Data output on ASCII command only
.
Ser 2
.
Serial Setup 2
__000
0U
Meter address with Custom
ASCII protocol
Select
0
thru
F
for addresses 1 thru 15.
Select
0.
thru
F.
(with decimal point) for
addresses 16 thru 31.
.
Ser 3
.
Serial Setup 3
_
00000U
RS485 half or full duplex
0
Full duplex
1
Half duplex
_0
0000U
Special start & stop char.
0
Standard continuous mode
1
Special start & stop characters
9