beautypg.com

FUTEK IPM500 (D500) Digital Display User Manual

Page 37

background image

0 0 0 300 baud

0 0 1 600 baud

0 1 0 1200 baud

0 1 1 2400 baud

1 0 0 4800 baud

1 0 1 9600 baud

1 1 0 19200 baud

0 Send unfiltered value

1 Send filtered value

30

Options

Do not use. This byte is determined by installed option boards.

2F

Filter

Bit 7 6 5 4 3 2 1 0 Approximate time constant:

0 0 0 No filtering

0 0 1 0.1 sec

0 1 0 0.2 sec
0 1 1 0.4 sec

1 0 0 0.8 sec

1 0 1 1.6 sec

1 1 0 3.2 sec

1 1 1 6.4 sec

0 Low adaptive threshold

1 High adaptive threshold

0 Display unfiltered input

1 Display filtered input

0 Peak value of unfiltered input

1 Peak value of filtered input

0 Adaptive filter

1 Conventional filter

2E

Setup

Bit 7 6 5 4 3 2 1 0 EXT IN A EXT IN B
0 0 0 0 Meter Reset Function Reset

0 0 1 0 Meter Reset Peak Display

0 0 1 1 Meter Reset External Gate

0 1 0 0 Function Reset Hold

0 1 0 1 Function Reset Peak Display

0 1 1 0 Function Reset External Gate

0 1 1 1 Hold Peak Display

1 0 0 0 Hold External Gate

1 0 0 1 Peak Display External Gate

1 0 1 0 Meter Reset Display Blank

1 0 1 1 Function Reset Display Blank

1 1 0 0 Hold Display Blank

1 1 0 1 Peak Display Display Blank

1 1 1 0 Display Blank External Gate

1 1 1 1 Display item#2 Display item#3

0 Scale2, Offset2 entered directly

1 Scale2, Offset2 using Coordinates of 2 points

0 Scale1, Offset1 entered directly

1 Scale1, Offset1 using Coordinates of 2 points

0 Blank leading zeros

1 Display leading zeros (0 = Blank leading zeros)
0 Zero the total at power on

1 Restore total at power-on

Dual Channel Signal Conditioner

Rate

Period

Total

Time Interval

41 A to B

Stopwatch

50 A to A

51 A to B

Phase

00 A,B

01 A only

02 Batch

03 A, Atot

05 A, Btot

0B A+B

0C A-B

0D AxB
0E A/B

0F A/B-1

10 A,B

11 A only

1B A+B

1C A-B

1D AxB

20 A,B

21 A only

24 A-B updwn

26 Burst

27 B, Arate

29 A,Bup/dwn

2A A, Binhibit

2B A+B
2C A-B

2D AxB

2E A/B

61 A to B

VF Converter

4-20 mA

0-1 mA

0-10V

81 A only

82 Batch

83 A, Atotal

88 Atot, A

8F 1/A

91 A only

92 Batch

93 A, Atotal

98 Atot, A

9F 1/A

A1 A only

A2 Batch

A3 A, Atotal

A8 Atot, A

AF 1/A

Quadrature

2B

Input Type

C0 Total

2C

Alarm Cnfg2

Bit 7 6 5 4 3 2 1 0 #Consecutive readings to Alarm

0 0 0 1

0 0 1 2

0 1 0 4

37