beautypg.com

Received signal, Bulletin f -9-pux2-s – Dwyer PUX2 User Manual

Page 30

background image

-28-

9.1. RECEIVED SIGNAL

Select Tab Page 2 on the MAINTE screen and the following screen appears.

Fig. 9-2 RECEIVED SIGNAL screen

x Select one from forward direction received wave, reverse direction received wave, forward direction filter, reverse

direction filter and calculation waveform.

x Click the [START] button to read in the specified cycle and click the [Stop] button to cancel the operation. Display is

updated with each read.

x Save the received waveform in the specified file by [SAVE As CSV] button. Trigger level is also displayed.
x Display the RAS information in the STATUS DISPLAY.

1. Startup is completed within 3 to 6 waves.
2. No peak (amplitude) fluctuation should be observed.

Otherwise air is mixed in.

Received
signal

Trigger level

Point

Bulletin F

-9-PUX2-S