Run 6, Totalizer (flow mode) 6, Password 6 – Metex 3200U Ultrasonic Flow & Level Monitor User Manual
Page 6: A qua m etrix

RUN
A scrolling display shows the units selected from the
UNITS/MODE
column, the mode of operation (
Range, Level
or
Flow
), the full scale value for the large numeric display, Totalizer
value (in
FLOW
mode), and air temperature at the sensor location.
ECHO CONFIDENCE
Echo Strength is automatically adjusted by the Ultrason in response
to operating conditions in the applications.
From RUN use
È to get to the Echo Confidence display,
EC
.
TOTALIZER
The Totalizer display will only be enabled when the Ultrason is calibrated in
FLOW
mode. From
RUN
use
È or Ç keys to display
Tot:
value. The Totalizer value is updated every 2 seconds with flow
volume > 1 litre (0.264 USG). The display will show up to 10 digits and then restart at 0 automatically.
Press
È or Ç keys to return to
RUN
.
The totalizer can be reset to zero by going
Æ to
SPECIAL FUNCTION
and
È to
Reset Tot?
.
RELAY STATUS DISPLAY
Press
È from
Tot:
to
RELAYS 1,2,3
.
Engerized relays will be displayed in reverse type
RELAYS
:
PASSWORD
The password (a number from
00
to
99
) prevents unauthorized access to the
CALIBRATION
menu.
From
RUN
(scrolling display) press
Æ to
PASSWORD
.
Press
Æ to place the cursor under the digits and È or Ç to change the
number and then
Æ to proceed to
CALIBRATION
.
Factory default Password is
00
. A new password can be stored by going
Æ to
SPECIAL FUNCTION
and
È to
New Password
.
AquaMetrix
Ultrason
Page 6
PASSWORD: 00
RUN
FLOW MODE ONLY
RELAYS:1 2 3 4 5 6
Tot:
EC
T: 9999999999
T: 0283168470
TOT:
0
TOT:
0
Units: ft
³, USG, IG, L
Units = m
³
Current
Echo Strength
Arrow
Minimum
Marker
EC
PASSWORD: 00
PASSWORD:
0
0
PASSWORD:
0
1