Special functions 15, A qua m etrix – Metex 3200U Ultrasonic Flow & Level Monitor User Manual
Page 15
SPECIAL FUNCTIONS
SLT32 V2.26
shows software version installed
View Codes? Yes
Select
Yes
and then
È to display an
instrument performance code. Faults detected
by the instrument are stored in the code and
will be requested by AquaMetrix if you call
for service or assistance.
Tag
Enter Instrument Tag number (0-9999)
Date MM DD/YYYY
(with optional Data Logger) Press
Æand Ç or
È to change Month, Day Year.
Time HH:MM:SS
(with optional Data Logger) Press
Æ and Ç or
È to change Hours, Minutes, Seconds
LOE TIME
Press
Æ and Ç or È to change the number of
seconds without receiving an echo before the
Ultrason displays
ECHO LOSS
, and Control
relays change state as calibrated under Relay
Parameters.
Factory default is 30 seconds and is
recommended for most applications,
Minimum is 10 seconds and maximum is 60
seconds.
Reset Tot? Yes
(
FLOW
mode only) Position the cursor under
Yes
to reset the totalizer. Store this selection
at the
Store? Yes
prompt.
Temp 25.0
°C
Indicates current temperature at the sensor
head. Press
Æ to
Min Temp
and
Æ to
Max
Temp
to display the minimum and maximum
temperatures the Ultrason has sensed. Press
Æ
to Display
°
C
°
F
. Position the cursor under
°
C
for celcuis display or
°
F
for Farenheight
display. Press
È to return to
Temp
.
ISB?: No
Yes
Select
Y
es only when the instrument is equipped with an optional Intrinsic
Safety Barrier for sensor mounting in hazardous locations. This setting
compensates for the electrical characteristics of an Intrinsic Safety Barrier.
AquaMetrix
Ultrason
Page 15
SPECIAL FUNCTION
SLT32
V 2.26
Tag
01
Date May 31/1999
Time
23:39:51
LOE Time
30 s
Reset
Tot?
Yes
Temp
20.3 C
º
Simul
0.00%
4mA ADJ
4.000
20mA ADJ
20.000
New Password: 00
Com 24 48 96 192
Store?
Yes
Min Temp
20.0 C
º
Max Temp
23.0 C
º
DISPLAY
F
C
º
º
V
Y
iew Codes?
es
1234567890123456
ISB?:
N
Yes
o