A qua m etrix – Metex 3200U Ultrasonic Flow & Level Monitor User Manual
Page 34
DATA LOGGING (Optional)
Setup
From
RUN STOP SETUP
press
Æ to
SETUP
and then
È to
Log Site ID 0
. Press Æ to position the
cursor under the numeral and
È or Ç to change the numerals. The “Site ID” number is retained with
data logging sessions to identify logs stored from different locations.
Formatted Data
Press
È from
Log Site ID
and press
È from
Formatted
.
“
Formatted
” data stores a summary of flow readings over a user-selectable time period. The
summary includes:
DATE and TIME
Interval TOTAL
Interval AVERAGE
Interval MAX FLOW
Interval MAX FLOW TIME
Interval MIN FLOW
Interval MIN FLOW TIME
From
Formatted
press
È to
Start MMM DD/YYYY
(eg:
Jan 01/2000
). Press
Æ to position the
cursor and then
È or Ç to set the Month, Day and Year that logging will Start. Press Æ to return to
Start
.
Press
È to
Start
(time) and
Æ to position the cursor under the time column
HH/MM/SS
(24 hour
clock in Hours/minutes/seconds, eg:
23:02:16
) and then
È or Ç to set the logging start Time. Press
Æ to return to
Start
.
Press
È to
Interval
and
Æ to the
Hrs
column. Press
È or Ç to select the flow logging interval.
Choose from:
24 Hrs
, or
12 Hrs
, or
8 Hrs
, or
4 Hrs
, or
1 Hrs
Press
Æ to return to
Interval
.
Press
È and the Ultrason will report
xxxxx Hrs Left
indicating
the amount of logging time available with your current set-up. You can also press
Ç to return to
previous menu items and make changes.
Press
È to
WrapAround Yes?
. Press
Æto
Yes?
and
È to enable the logging wrap function. In
WrapAround mode the oldest data will be overwritten by the newest. If WrapAround is not enabled the
logger will stop when its memory becomes full.
Press
È to
Reset Log? Yes
. Press
Æ to
Yes
and then
È to reset the Log and erase all previous
sessions and stored values. Or press
È from
Reset Log?
to retain existing data in the Log. The
Ultrason will display “
xxxxx Hrs/Days Left
.
AquaMetrix
Ultrason
Page 34