A qua m etrix – Metex 3200U Ultrasonic Flow & Level Monitor User Manual

Page 37

background image

Press

È to

Interval

and

Æ to the

Sec/Min

column. Press

È or Ç to set the logging time interval.

Choose:

30 Sec

10 Sec

5 Sec

2 Sec

1 Sec

30 Min

10 Min

5 Min

2 Min

1 Min

Press

Æ to return to

Interval

and press

È to

Reset Log?

To erase all existing data in the log press

Æ to

Yes

and

È. To keep existing data in the Log press È from

Reset Log?

If you have made

changes to the Start Date, Time or Interval, the data logger will automatically start a new “session”. The
Ultrason will display “

xxxxx Hrs/Days Left

”.

From the

xxxxx Hrs Left

display press

È to

Store? Yes

. Press

Æ to

Yes

and then

È to save

your Data Logging setup, or press

È from

Store?

to cancel changes made above and exit without

storing changes.

From the Data Logging

Store? Yes

prompt the menu will return to

RUN STOP SETUP

. Press

Æ to

position the cursor under

RUN

and press

È to activate the Data Logger to start. The Ultrason will

display

SESSION NO. x

. Press

È to return to

DATA LOGGING

.

Logging "Sessions"

Each time you select

STOP

in the

DATA LOGGING

menu, the Data Logger stores the current data in

memory as a "

SESSION NO

" automatically numbered from "

1

" to "

10

". If you resume logging by

selecting

RUN

, the Data Logger will report that a new logging session is started and titled "

SESSION

NO xx

". When you download the logger files to your PC using AquaMetrix Logger software, each

Session will open as a separate graph/table titled "AquaMetrix Data Log xx".

Important:
If you

STORE

instrument calibration changes under the

UNITS/MODE

or

CALIBRATON

menus,

STOP

the data logger and select

RUN

again to start a new logging Session with your new calibration values.

AquaMetrix

Ultrason

Page 37