A qua m etrix – Metex 3200U Ultrasonic Flow & Level Monitor User Manual
Page 37
Press
È to
Interval
and
Æ to the
Sec/Min
column. Press
È or Ç to set the logging time interval.
Choose:
30 Sec
10 Sec
5 Sec
2 Sec
1 Sec
30 Min
10 Min
5 Min
2 Min
1 Min
Press
Æ to return to
Interval
and press
È to
Reset Log?
To erase all existing data in the log press
Æ to
Yes
and
È. To keep existing data in the Log press È from
Reset Log?
If you have made
changes to the Start Date, Time or Interval, the data logger will automatically start a new “session”. The
Ultrason will display “
xxxxx Hrs/Days Left
”.
From the
xxxxx Hrs Left
display press
È to
Store? Yes
. Press
Æ to
Yes
and then
È to save
your Data Logging setup, or press
È from
Store?
to cancel changes made above and exit without
storing changes.
From the Data Logging
Store? Yes
prompt the menu will return to
RUN STOP SETUP
. Press
Æ to
position the cursor under
RUN
and press
È to activate the Data Logger to start. The Ultrason will
display
SESSION NO. x
. Press
È to return to
DATA LOGGING
.
Logging "Sessions"
Each time you select
STOP
in the
DATA LOGGING
menu, the Data Logger stores the current data in
memory as a "
SESSION NO
" automatically numbered from "
1
" to "
10
". If you resume logging by
selecting
RUN
, the Data Logger will report that a new logging session is started and titled "
SESSION
NO xx
". When you download the logger files to your PC using AquaMetrix Logger software, each
Session will open as a separate graph/table titled "AquaMetrix Data Log xx".
Important:
If you
STORE
instrument calibration changes under the
UNITS/MODE
or
CALIBRATON
menus,
STOP
the data logger and select
RUN
again to start a new logging Session with your new calibration values.
AquaMetrix
Ultrason
Page 37