ERROR/WARNING MESSAGES

! LOSING ECHO !

No valid echoes for half

LOE TIME

(

SPECIAL FUNCTION

menu).

"ECHO LOSS"

(flashing)

No valid echoes received within the

LOE TIME

setting. See FIELD

TROUBLESHOOTING (F). In

Level

mode the Ultrason will hold the

display and outputs at the last reading until a new echo is received. In

FLOW

mode it will display

0.00

and totalization will stop until an echo

is received.

Er: ILLEGAL MinRg

The value entered for

MinRg

(Minimum Range) is less than 8"(20.3 cm).

MinRg

must be greater than or equal to 8" (20.3 cm) for US12-1A1A

sensors. Minimum Range for US2-1A1A sensors must be

³ 12" / 30.5

cm and for optional US2-2A1B sensor must be greater than or equal to
16" (40.6 cm).

Er: ILLEGAL UNITS

Your choice of UNITS exceeds a maximum value of 9,999,999. Use US
MGD, Imp MGD or m3/d so that UNITS will be 9,999,999 or less.

Err:ILLEGAL SPAN

1. The value entered for

MaxRg

is the same or less than

MinRg

+2".

Maximum range must be greater than minimum range and less than
102 ft. (31.09 m). (Note: to invert the scale and outputs, choose

Level

instead of

Range

selection in the

UNITS/MODE

menu.)

2. For Rectangular Weirs with End Contractions, the SPAN (

MaxRg -

MinRg

) must be less than 1/2 the weir size (eg: 12" crest width,

SPAN must be 6" or less).

- or -

Your choice of Units exceeds 9,999,999. Use

USMG/d

,

IMG/d

or

m3/d

so that Units will be 9,999,999 or less

ERR: ILLEGAL

SETPOINTS

ON or OFF setpoint is <

MinRg

or >

MaxRg

NORMAL / REVERSED

SET POINTS MIXED!

Relays have been selected for Pump Alternation and have been calibrated
in opposite modes. Any alternating relays must be calibrated in the same
mode:
1. with ON setpoint > OFF setpoint, or
2. with OFF setpoint > ON setpoint

TEMP * 23.5 C

* Indicates Temperature Compensation fault. Check sensor connections.

ECHO TOO CLOSE

Indicates that the target is less than

MinRg

distance from the sensor

(too close to the sensor)

!! SENSOR OPEN !

Instrument has detected sensor connection/cable Open.

! SENSOR SHORTED !

Instrument has detected sensor connection/cable Shorted.