Calibration worksheet - flow mode 47, A qua m etrix – Metex 3200U Ultrasonic Flow & Level Monitor User Manual
Page 47

CALIBRATION WORKSHEET - Flow Mode
for Open Channel Flow through Flume and Weirs
MAX RANGE
(Ultrason READS)
MAX HEAD
(Height at MAX Flow)
A
-
B
= MIN RANGE
______ -
______ =
__________
AquaMetrix
Ultrason
Page 47
A=______________
CALIBRATION
MODE
MIN RANGE
MAX RANGE
STORE?
YES
Program MIN RANGE then
MAX RANGE via:
A
B
SENSOR
MIN RANGE
MUST BE
12 in. (305 mm)
OR MORE
UNITS
SELECTION
ft in m cm %
RANGE
STORE?
YES
Program Ultrason to show
RANGE via:
B=______________