beautypg.com

A qua m etrix, Ultrason page 31, Am qua etrix – Metex 3200U Ultrasonic Flow & Level Monitor User Manual

Page 31

background image

SENSOR INTRINSIC SAFETY - OPTION ISB
Ultrason
Sensors (with built-in temperature compensation)

When connected through an Intrinsic Safety Barrier, the AquaMetrix Ultrason sensors are CSA certified
for installation in a hazardous location rated:

Class I, Groups C,D
Class II, Groups E, F and G
Class III

Intrinsic Safety Barriers may be ordered with the AquaMetrix instrument and are supplied mounted in
the AquaMetrix instrument enclosure. Replacement barrier fuses (Part No. ISB-011239) may be
purchased separately. Barriers must be installed in the sensor cable between the safe and hazardous
locations, and must be mounted in either the safe or Div. 2 area. Barriers may be plate, busbar or rail
mounted.

Intrinsic Safety Barrier Specifications: Certified, rated 17.5V max, 95 ohms min. (Recommended: Stahl
Model 9001/02-175-200-10).

AquaMetrix

Ultrason

Page 31

T.DUCER

N.O C N.C N.O C N.C N.O C N.C

RELAY 1

RELAY 2

RELAY 3

L N G

AC

RX PU GND TX

RS 232

N.O C N.C N.O C N.C

RELAY 4

RELAY 5

N.O C N.C

RELAY 6

4 - 20

OPTIONAL

3

4

1

2

NOTE:

MUST BE

BARRIER-EQUIPPED UNITS
ARE FACTORY-WIRED WITH
GROUND THROUGH THE
INSTRUMENT CHASSIS.
POWER INPUT GROUND

CONNECTED TO A

GOOD GROUND (<1 Ohm) WITH
A 12 AWG CONDUCTOR

REPLACEMENT FUSE ASSEMBLY
ORDER PART NO. ISB-011239 (160mA)

NON-HAZARDOUS LOCATION

HAZARDOUS LOCATION
CLASS I, GROUPS C,D
CLASS II, GROUPS E,F,G
CLASS III

INTRINSIC SAFETY BARRIER

SENSOR

A

M

qua etrix